Chapter IV Local Operation
83
ARRAY
then select
Value
option via TAB key; set the value to 200.00 via
and
keys or via the
knob; then select
Time
option via TAB key; set the value to 1.0000 via
and
keys or
via the knob;
Step 13:
Press dynamic
key to return to the list operation menu;
Step 14:
Press dynamic
key to start the current list operation program and it will stop
automatically when the operation is finished;
Step 15:
During the list operation, press dynamic
key to stop the ruing list operation.
The corresponding SCPI commands of the above procedures should be as follows:
LIST
:
NUMB
0
;
Call the sequence numbered 0
LIST
:
MEMO
“Test0”
;
Name the sequence Test.
LIST
:
EDIT
0, CCL,1A,1S
LIST
:
EDIT
1, CCH,20A,1S
LIST
:
EDIT
2, CVH,10V,1S
LIST
:
EDIT
3, CRL,1R,1S
LIST
:
EDIT
4, CP,200W,1S
;
The above five commands are the sequence steps
LIST
:
COUN
10
;
Cyclic time
LIST
:
CHA
OFF
;
Close the chained sequence of the list
LIST
ON
;
Start the list operation
LIST
OFF
;
Stop the list operation
4.8.2 Starting/Stopping the List
Example:
Start the No. 0 sequence and stop the active list operation.
Step 1:
Press dynamic
key to enter into list operation mode at initial interface;
Step 2:
Set the NO to 0 via the keypad or via the knob and press
key to confirm (The recall
number should be the sequence number in EEPROM.);
Step 3:
Press dynamic
key to start the current list operation
Step 4:
Press dynamic
key to stop the list operation;
Содержание 3750A
Страница 1: ...Contents USER S MANUAL 375X SERIES ELECTRONIC LOAD ...
Страница 63: ...Chapter III Installation 59 ARRAY ...
Страница 114: ...Chapter V Remote Programming Operation 110 ARRAY Frame Diagram Fig 1 ...
Страница 115: ...Chapter V Remote Programming Operation 111 ARRAY ...
Страница 116: ...Chapter V Remote Programming Operation 112 ARRAY Fig 2 ...
Страница 120: ......