![Array electronic 3750A Скачать руководство пользователя страница 69](http://html1.mh-extra.com/html/array-electronic/3750a/3750a_manual_2975640069.webp)
Chapter IV Local Operation
65
ARRAY
the number keypad; move the cursor to the digit position via the
key, then enter number 0
via the knob or the number keypad; then press
key to confirm;
Step 5:
Switch to voltage limit option via
and
keys;
Step 6:
Move the cursor to the hundred position via the
key, then enter number 1 via the
knob or the number keypad; Move the cursor to the ten position via the
key, then enter
number 0 via the knob or the number keypad; move the cursor to the digit position via the
key, then enter number 0 via the knob or the number keypad; press
key to confirm;
Step 7:
Switch to voltage rising rate option via
and
keys;
Step 8:
Move the cursor to the ten position via the
key, then enter number 2 via the knob or
the number keypad; move the cursor to the digit position via the
key, then enter number 0
via the knob or the number keypad; press
key to confirm;
Step 9:
Switch to voltage falling rate option via
and
keys;
Step 10:
Move the cursor to the ten position via the
key, then enter number 2 via the knob
or the number keypad; move the cursor to the digit position via the
key, then enter number
4 via the knob or the number keypad; press
key to confirm;
Step 11:
Press
key to open the load.
Notes: The knob changes the set value immediately. While using the number keypad, the
key is needed to press to confirm. Otherwise the operation will be invalid. When the
load is open, the set value will be valid in the input terminal immediately.
The corresponding SCPI commands of the above procedures should be as follows:
MODE
CVH
;
Set the mode
VOLT
60
;
Set the voltage level
VOLT:SLEW:POS 20
;
Set the voltage rising rate 20V/us
VOLT:SLEW:NEG 24
;
Set the voltage falling rate 24V/us
VOLT:STAR 20
;
Set the VON voltage 20V
Содержание 3750A
Страница 1: ...Contents USER S MANUAL 375X SERIES ELECTRONIC LOAD ...
Страница 63: ...Chapter III Installation 59 ARRAY ...
Страница 114: ...Chapter V Remote Programming Operation 110 ARRAY Frame Diagram Fig 1 ...
Страница 115: ...Chapter V Remote Programming Operation 111 ARRAY ...
Страница 116: ...Chapter V Remote Programming Operation 112 ARRAY Fig 2 ...
Страница 120: ......