Chapter IV Local Operation
72
ARRAY
The corresponding SCPI commands of the above procedures should be as follows:
MODE
CRL+CV
;
Set the mode
RES
1.5
;
Set the resistance level
VOLT:ADD:LIM 20
;
Set the voltage limit level
INP
ON
;
Open the load
4.6.3 CP+CV Mode
Example :
Take 3751A as an example, in CP mode, set the power to 1000W, the voltage limit to
30V. Turn on the load. The procedures are shown as below:
Step 1:
Switch to
mode by pressing
and
key simultaneously;
Step 2:
Switch to parameter setting menu via
and
keys;
Step 3:
Enter the current level 1000 by the number keypad, or via knob with
and
keys;
Step 4:
Press
key to confirm;
Step 5:
Switch to voltage limit option via
and
keys;
Step 6:
Enter the voltage level 30.000 by the number keypad, or via knob with
and
keys;
Step 7:
Press
key to confirm;
Step 8:
Press
key to open the load.
The corresponding SCPI commands of the above procedures should be as follows:
MODE
CP+CV
;
Set the mode
POW
1000
;
Set the power level
VOLT:ADD:LIM 30
;
Set the voltage limit level
INP
ON
;
Open the load