Chapter III Installation
53
ARRAY
and reduce the measurement error, a thick cable is suggested to use in a large current test.
3. Remote Voltage Test Terminal
There are two remote voltage test terminals (SENSE+, SENSE-) one the rear panel, for the remote
voltage test to prevent the affect on the voltage test of the voltage drop of the load input wiring
and to improve the accuracy of the voltage measurement. ARRAY 375XA Electronic Load can
detect the remote voltage input signal automatically, and it is unnecessary to modify the parameter
settings or change the hardware wiring in remote voltage test.
4. Communication Interfaces
RS-232 interface
The load provides an RS-232 interface, which is a standard DB9 pin connector, and uses the DTR
and DSR signals for flow control. The RS-232 Connector Pins is as below:
Pin Input/Output Description
1 - No connection
2 Input RXD, receive data
3 Output TXD, transmit data
4 - No connection
5 Common GND, signal ground
6 - No connection
7 - No connection
8 - No connection
9 - No connection
The interface parameters can be set in the Com Port Configuration Menu as in the following
diagram, and you can use SCPI language for programming to realize the communication with the
load.
Содержание 3750A
Страница 1: ...Contents USER S MANUAL 375X SERIES ELECTRONIC LOAD ...
Страница 63: ...Chapter III Installation 59 ARRAY ...
Страница 114: ...Chapter V Remote Programming Operation 110 ARRAY Frame Diagram Fig 1 ...
Страница 115: ...Chapter V Remote Programming Operation 111 ARRAY ...
Страница 116: ...Chapter V Remote Programming Operation 112 ARRAY Fig 2 ...
Страница 120: ......