Chapter V Remote Programming Operation
100
ARRAY
5.7 CV Mode Example
Example:
Preset the trigger voltage to 5V, and then select the external trigger source.
l "INP OFF"
;
Turn off the load’s input
2 "MODE CV"
;
Select the CV Mode
3 "VOLT:TRIG 5"
;
Preset the trigger voltage to 5V
4 "TRIG:SOUR EXT"
;
Select the external input as the trigger source
5"INP ON"
;
Turn on the load’s input
In this example, when the external trigger signal is received, the input voltage will be set to 5V.
5.8 CR Mode Example
Example:
Set the current protection limit to 10A, the time delay to 10s, the resistance level to
10
Ω
and return back the computed power level.
l. "INP OFF"
;
Turn off the load’s input
2. "MODE CRL"
;
Select the CRL Mode
3. "CURR:PROT 10"
;
Set the current protection level to 10A
4. "CURR:PROT:DEL 10"
;
Set the current protection delay time to 10s
5. "CURR:PROT:STAT ON"
;
Active the current protection
6. "RES 10"
;
Set the resistance level to 10
Ω
7. "INP ON"
;
Turn on the load’s input
8. "MEAS:POW?"
;
Measure the input power level
5.9 Continuous Transient Operation Example
Example:
Set the voltage transient high level, voltage transient low level, time for transient rising
edge, time for transient falling edge, high level time, low level time and parameters for transient
operations.
1 "INP OFF"
;
Turn off the load’s input
2 "MODE CVL"
;
Select the CVL Mode
3 "SYST:STAT TRAN"
;
Open the transient
Содержание 3750A
Страница 1: ...Contents USER S MANUAL 375X SERIES ELECTRONIC LOAD ...
Страница 63: ...Chapter III Installation 59 ARRAY ...
Страница 114: ...Chapter V Remote Programming Operation 110 ARRAY Frame Diagram Fig 1 ...
Страница 115: ...Chapter V Remote Programming Operation 111 ARRAY ...
Страница 116: ...Chapter V Remote Programming Operation 112 ARRAY Fig 2 ...
Страница 120: ......