Chapter III Installation
57
ARRAY
The diagram above shows how changes in current can affect voltage.
Load line inductance can be reduced in two ways:
1.
Ensure load wires being as short as possible;
2.
Twist the positive and negative load wires together.
Current change can be limited by limiting the slew rate or response speed when switching in CR
and CC mode.
Load Wire Connections
The input terminal of the load is on the rear panel. Strictly follow the procedure below for all the
load connections to ensure the safety of the personal and the instrument. Ensure the polarity is
correct before proceeding with any connection. Ensure the maximum input voltage is less than
240V.