Chapter II Functions & Features
31
ARRAY
2.3.7 CP+CV Test
When CP
+
CV mode is enabled, if the input voltage is more than the limit voltage, no matter how
the input current is changed ,the Eload will always enter the voltage at the limit of the voltage.
If
the input voltage is less than the limit voltage, no matter how the input voltage changes, the power
of the Eload will be always at the setting power.
2.4 Transient Operation
In transient operation mode, the load can continuously toggle between the two levels: the transient
high level (LevelH) and the transient low level (LevelL), which can be applied to test the dynamic
features of the power source. The transient operation can be executed in CC, CV, CR and CP mode
and there are three operation statuses: Continuous, Pulse and Toggle. Please make sure that the
List Operation function has been turned off before the transient operation.
The transient operation parameters include: transient low level (LevelL), transient high level