Recommended separation distances between portable and mobile RF communications equip-
ment and the
Alenti
.
Alenti
is intended for use in an electromagnetic environment in which radiated RF disturbances
are controlled.
The customer or the user of
Alenti
can help prevent electromagnetic interference by maintain-
ing a minimum distance between portable and mobile RF communications equipment (trans-
mitters) and the
Alenti
by maintaining a distance of not less than 1 meter to any part of
Alenti
,
including cables, if the transmitters output power rating exceeds 1 W.
Guidance and manufacturer’s declaration - electromagnetic immunity
Emissions test
IEC 60601-1-2
test level
Compliance
Electromagnetic environment -
guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
6 kV contact
8 kV air
6 kV contact
8 kV air
Floors should be wood, concrete or
ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30%.
Radiated RF
IEC 61000-4-3
3 Vrms
80 MHz to
2,5 GHz
10 Vrms
80 MHz to
2,5 GHz
Portable and mobile RF communica-
tions equipment should be used no
closer to any part of
Alenti
, including
cables, than 1 meter if the transmit-
ters output power rating exceeds 1
W. Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey,
A) should be less than the compli-
ance level in each frequency range.
B) Interference may occur in the
vicinity of equipment marked with
the following symbol:
EN ISO
10535:2006
10 v/m
10 v/m
a) Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV
broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic
environment due to fixed RF transmitters, an electromagnetic site survey should be consid-
ered. If the measured field strength in the location in which
Alenti
is used exceeds the appli-
cable RF compliance level above, the
Alenti
should be observed to verify normal operation.
If abnormal performance is observed, additional measures may be necessary.
b) Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 1 V/m.
Содержание Alenti
Страница 1: ...with people in mind 0086 Alenti 04 CD 05_9US CA June 2012 INSTRUCTIONS FOR USE...
Страница 9: ...9 8 11 2 1 3 10 12 9 13 14 15 16 17 19 18 7 5 4 6 20...
Страница 13: ...13 B A 1 2 3 4 5 A...
Страница 15: ...15 A 1 2 3 4 5 6 7...
Страница 17: ...17 1 2 3 4 5 6 7 8...
Страница 19: ...19 1 2...
Страница 21: ...21 2 1...
Страница 23: ...23 1 2...
Страница 25: ...25 1 4 3 2...
Страница 27: ...27 1 2...
Страница 29: ...29 1 2 3...
Страница 31: ...31 1 2 3...
Страница 33: ...33 1 2 3 4...
Страница 35: ...35 1 2 3 4...
Страница 41: ...41 1 2 3...
Страница 43: ...43 1 2 3 4...
Страница 45: ...45 Intentionally left blank...
Страница 49: ...49 1 2 3 4 5 6...
Страница 61: ...61 Intentionally left blank...
Страница 62: ...62 Intentionally left blank...