48
Caregiver Obligations
Caregiver obligations shall be carried out by
personnel with sufficient
Alenti
knowledge
following the instructions in this
IFU
.
Every day
•
Disinfect. Alenti
has to be disinfected
immediately after every use.
ArjoHuntleigh disinfectant agents should
be used in recommended concentrations.
See section,
Disinfection instructions.
Every week
•
Check the mechanical attachments
by
placing a foot on the chassis and pull the
seat upwards with both hands.
(See Fig. 1)
No gaps between the chassis and seat are
allowed.
•
Visually check the safety belt and leg
straps.
Check the complete length for
fraying, cuts, crack, tears, that no inner
fabric is showing and that no other dam-
age is visible.
(See Fig. 2)
If the safety belt is found damaged in any
way, replace it.
•
Check hand control and cable:
Visu-
ally check the condition of the hand con-
trol and its cable.
(See Fig. 3)
If found cut or damaged, replace it.
•
Check that the castors are properly
fixed
and are rolling and swivelling
freely.
(See Fig. 4)
Clean with water (the function can be
affected by soap, hair, dust and chemicals
from floor cleaning).
•
Check the central brake function and
brake pads
by activating the brakes and
try to move the
Alenti
.
(See Fig. 5)
It must be impossible to move without
lifting the
Alenti
from the floor.
•
Visually check the condition of the bat-
tery charger
, its cable and the battery.
(See Fig. 6)
Withdraw and replace immediately if
damaged.
•
Visually check all exposed parts
, espe-
cially where personal contact is made by
either the resident or caregiver. Make
sure no cracks or sharp edges have devel-
oped that could cause the resident or car-
egiver injury or that has become
unhygienic. Replace damaged parts.
•
Check for corrosion and damage:
if
there is any visible corrosion or damage
on parts, contact your local Arjo-
Huntleigh representative for support.
•
Perform a functionality test:
Test up and down motion by pressing the
up and down button on the hand control
and control panel. Also test the Emer-
gency High/Low and brakes.
Yearly
Alenti
has to be serviced according to the
Preventive Maintenance Schedule (Quali-
fied Personnel Action/Check).
Содержание Alenti
Страница 1: ...with people in mind 0086 Alenti 04 CD 05_9US CA June 2012 INSTRUCTIONS FOR USE...
Страница 9: ...9 8 11 2 1 3 10 12 9 13 14 15 16 17 19 18 7 5 4 6 20...
Страница 13: ...13 B A 1 2 3 4 5 A...
Страница 15: ...15 A 1 2 3 4 5 6 7...
Страница 17: ...17 1 2 3 4 5 6 7 8...
Страница 19: ...19 1 2...
Страница 21: ...21 2 1...
Страница 23: ...23 1 2...
Страница 25: ...25 1 4 3 2...
Страница 27: ...27 1 2...
Страница 29: ...29 1 2 3...
Страница 31: ...31 1 2 3...
Страница 33: ...33 1 2 3 4...
Страница 35: ...35 1 2 3 4...
Страница 41: ...41 1 2 3...
Страница 43: ...43 1 2 3 4...
Страница 45: ...45 Intentionally left blank...
Страница 49: ...49 1 2 3 4 5 6...
Страница 61: ...61 Intentionally left blank...
Страница 62: ...62 Intentionally left blank...