5
Instructions For Use
EN
Team3 series monitors are not intended for use with patients fitted with cardiac pacemakers.
Do not use with electrosurgical devices.
Team3 can be isolated from the AC mains supply by removing the IEC mains inlet connector. Ensure
that this is fully accessible at all times.
Team3 is a Class 1 product that relies for safety on its protective earth. Ensure it is connected to a
suitably earthed AC mains supply.
Do not use in the home environment.
Do not use the Team3 in vehicles or in aircraft.
If this product is connected to another item of electrical equaipment, ensure that the system is fully
compliant with IEC60601-1:2005.
This product contains sensitive electronics, therefore, strong radio frequency fields could possibly
interfere with it. This may be indicated by unusual sounds from the loudspeaker. We recommend that
the source of interference is identified and eliminated.
Do not expose to excessive heat, including prolonged exposure to sunlight.
This equipment must not be modified.
This equipment is for use only by suitably qualified healthcare practitioners.
When configuring the system, consider and minimise the risk of persons tripping over cables.
Do not use during magnetic resonance imaging (MRI) scanning.
Do not use if there is any damage to the unit or its accessories.
The use of the Team3 is restricted to one patient at a time.
The risk of cyber attack on the fetal monitor is negligible. No special means are required to secure the
device or its updates.
The emissions characteristics of this equipment make it suitable for use in individual areas and
hospitals (CISPR 11 Class A). If it is used in a residential environment, (for which CISRR 11 Class
B is normally required), this equipment might not offer adequate protection to radio-frequency
communication services. The user might need to take mitigation measures, such as re-locating or re-
orienting the equipment.