11. Trouble Shooting
This section gives some of the more common problems encountered during use together with possible causes.
If the problem cannot be located after consulting the table in this section, the Team3 should be switched off,
disconnected from mains power source and a qualified technician should be consulted.
Before attempting trouble-shooting, verify that the power cable is properly connected to both the Team3 and
the mains power source.
11.1 Essential Performance
If you lose essential performance, refer to the table on page 98, Manufacturer's Performance Criteria
Specification.
11.2 FHR
SYMPTOM
POSSIBLE CAUSE / REMEDY
No FHR signal displayed
Check Team3 is switched on.
Check the FHR transducer is connected.
High % signal loss
Check transducer placement.
Check the transducer is not damaged.
Consider switching from Ultrasound to FECG.
No FHR trace printed
Check Print button has been pressed.
Check there is paper in the paper tray.
Check paper tray is fully pushed in.
Only one trace (twins) OR Traces
superimposed (twins)
Check ‘Printer Offsets’ in Printer setup. See Section 5.3.7
No beep when button pressed.
Check sound setting. See section 5.3.7.
Alarm not working
Alarm may be turned off. See Section 5.3.7
Team3 not printing/recording.
11.3 Fetal event marker
SYMPTOM
POSSIBLE CAUSE / REMEDY
No mark appears on the trace when the
mother presses the event marker.
Check event marker is connected
Check enough time has elapsed since button last pressed.
Team3 does not beep when the mother
presses the event marker
Check event marker is connected
Check sound setting. See section 5.3.7.
11.4 Maternal blood pressure
SYMPTOM
POSSIBLE CAUSE / REMEDY
No reading reported.
Check cuff and hose, then try another measurement.