79
Instructions For Use
EN
Guidance and Manufacturer’s declaration - electromagnetic immunity
The Team3 is intended for use in the electromagnetic environment specified below. The customer or the user
of the Team3 should assure that it is used in such an environment.
Immunity Test
IEC 60601 test
level
Compliance level Electromagnetic Environment -
guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30%.
Electrical fast
transient burst
IEC 61000-4-4
± 2 kV for power
supply lines
± 1 kV for input/
output lines
± 2 kV for power
supply lines
± 1 kV for input/
output lines
Mains power quality should be that
of a typical commercial or hospital
environment.
Surge
IEC 61000-4-5
± 1 kV line(s) to
line(s)
± 2 kV line(s) to
earth
± 1 kV line(s) to
line(s)
± 2 kV line(s) to
earth
Mains power quality should be that
of a typical commercial or hospital
environment.
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines
IEC 61000-4-11
<5 % Ur (>95 %
dip in Ur )
for 0,5 cycles
40 % Ur (60 % dip
in Ur )
for 5 cycles
70 % Ur (30 % dip
in Ur )
for 25 cycles
<5 % Ur (>95 %
dip in Ur )
for 5 s
<5 % Ur (>95 % dip
in Ur )
for 0,5 cycles
40 % Ur (60 % dip
in Ur )
for 5 cycles
70 % Ur (30 % dip
in Ur )
for 25 cycles
<5 % Ur (>95 % dip
in Ur )
for 5 s
Mains power quality should be that
of a typical commercial of hospital
environment. If the user of the Team3
requires continued operation during power
mains interruptions, it is recommended
that the Team3 is powered from an
uninterruptible power supply or battery,
by specifying the battery option at time of
purchase.
Power frequency
(50/60Hz) magnetic
field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields should
be at levels characteristic of a typical
location in a typical commercial or hospital
environment.
NOTE Ur is the a.c. mains voltage prior to the application of the test level.