Guidance and Manufacturer’s declaration - electromagnetic immunity
The Team3 is intended for use in the electromagnetic environment specified below. The customer or the user
of the Team3 should assure that it is used in such an environment.
Immunity Test IEC 60601 test level
Compliance
level
Electromagnetic Environment - guidance
Proximity
Immunity
Compliance
Information
a The ISM (industrial, scientific and medical)
bands between 150 kHz and 80 MHz are 6,765
MHz, to 6,795 MHz; 13,553 MHz to 13,567 MHz;
26,957 MHz to 27,283 MHz; and 40,66 MHz to
40,70 MHz
Conducted RF
IEC 61000-4-6
3 Vrms
150kHz to 80MHz
outside ISM bandsa
6 Vrms
150 kHz to 80 MHz in
ISM and amateur radio
bands
3V
Radiated RF
IEC 61000-4-3
3 V/m
80MHz to 2.5GHz
3V/m
80MHz to 800MHz
800MHz to 2.5GHz
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey, b
should be less than the compliance level in each
frequency range c.
Interference may occur in the vicinity of the
equipment marked with the following symbol:
NOTE 1 At 80MHz and 800MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
b Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones
and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted
theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an
electromagnetic site survey should be considered. If the measured field strength in the location in which the
Team3 is used exceeds the applicable RF compliance level above, the Team3 should be observed to verify
normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-
orienting or relocating the Team3.
c Over the frequency range 150kHz to 80kHz, field strengths should be less than 3V/m.