70
Instructions For Use
EN
Maternal blood pressure
Method
Oscillometric
Pressure Range
0-300mmHg
Measurement ranges
Systolic 25-280 mmHg
Diastolic 10-220 mmHg
Pulse 30-240 bpm
Accuracy
Measurement
during deflation
Measurement during
inflation, IMT
Required according to
international standards
Pressure transducer
accuracy
±1 mmHg
±1 mmHg
max. ±3 mmHg
Measurement accuracy
mean deviation
<1.7 mmHg
<1.19 mmHg
max. ±5 mmHg
Measurement accuracy
standard deviation
<5.6 mmHg
<3.48 mmHg
max. 8 mmHg
Modes
Manual or automatic
User-selectable interval in Auto Mode:
1, 2, 3, 5, 10, 15, 20, 30, 45, 60, 90 or 120 minutes
Record / display
On-screen display and printed record of:
●
Systolic blood pressure
●
Diastolic blood pressure
●
Pulse rate
●
Mean arterial pressure
Alarms
Systolic High: 100-180mmHg
Systolic Low: 50-150mmHg
Diastolic High: 70-130mmHg
Diastolic Low: 40-120mmHg
Safety
Type B protection
Hardware and software controls to limit:
●
Inflation (max. 300 mmHg)
●
Measurement time (max. 90 s)
Standards
EN 1060-3
EN 1060-4
EN ISO 81060-2
EN 80601-2-30
ANSI/AAMI SP-10