68
Instructions For Use
EN
14. Specifications
14.1 Equipment Classification
Protection against electric
shock.
Class 1
Applied Parts
Type CF - Ultrasound Probes /TOCO/ FECG
Type BF - Maternal NIBP/fetal event marker
Mode of operation.
Continuous
Degree of protection against
harmful ingress of particles
and/or water.
Main Unit : IP30 when Fixed or Stationary.
IP32 with protective cover used
when moving product
Ultrasound and Toco: IPX7
Other transducers : Not protected
Suitability for use in an
oxygen rich environment.
Not suitable
14.2 General
Rated Supply Voltage
100-230V AC
Fuse Type
2 x T3.15AH 250V
Power Input
50-60Hz 8-133VA
Battery (optional)
14.4V Lithium Ion Battery Pack
Real time clock battery
Panasonic CR2032/BN 3V lithium
Size
Width 318mm, Height 230mm, Depth 237mm
Weight
5.7Kg (with Printer)
Service Life
7 years
14.3 Environmental
Operating
Storage
+10°C to +40°C
Temperature range
-10°C to +40°C
10% to 90% (non condensing)
Relative Humidity
93% maximum
86 kPa to 106 kPa
Pressure
86 kPa to 106 kPa