1. Safety
We recommend that exposure to ultrasound should be kept As Low As Reasonably
Achievable - (ALARA guidelines). This is considered to be good practice and should be
observed at all times.
Team3 provides just one indicator of fetal condition. This should be assessed as part of an
holistic approach to obstetric care together with other factors. A complete assessment must
be made before taking appropriate action. If there is any doubt concerning the accuracy of
any measurement, an alternative method should be used.
Symbols
General Warning
Refer to Instructions for Use
Caution: Federal law restricts this device to sale by, or on the order of a licensed healthcare
practitioner.
Rx Only
Attention, consult accompanying documents / Instructions for Use
1.1 Warnings
Do not use in the presence of flammable gases or in oxygen rich environments.
Do not sterilise the product or its accessories. The product will be damaged, and there is a risk of
patient and user harm.
Keep dry, do not immerse Team3 in liquid. Ultrasound and Toco transducers are IPX7 rated. Team3
with wired transducers is not intended for use in water birth situations.
Always fit Protective cover to protect against fluid ingress when moving Team3 by hand or on a trolley.
Do not use in the sterile field unless additional barrier precautions are taken.
Use only recommended accessories listed in this manual.
Do not dispose of batteries in fire as this can cause them to explode.
The optional Lithium battery pack is a service replaceable item. Replacement by inadequately trained
personnel could result in a hazard.
Do not use with defibrillators. Ensure that all Team3 leads and applied parts are removed from the
patient before applying Defibrillation.