8
GA 7000C-E-0313
© 2013 by arcomed ag, Switzerland
2
Specifications
Classification
IIb
Software revision:
4.xxxx
Flow rate range
0
– 1200 ml/h
Flow rate increments
0,1 ml/h, optional 0,01 ml/h
Volume range
0
– 9999 ml
Volume increments
0,1 ml, optional 0,01 ml
Deviation of flow rate with approved
Infusion set (e.g. 3101 P - PVC)
typ. +/- 5% (PVC tubing)
Deviation of flow rate with approved
Infusion set (e.g. 8101 P - Silicone)
typ. +/- 3% (Silicon segment)
Over infusion in case of error (Electr./Mech. error)
max 1,5ml
Keep Open Rate „KVO“
3 ml/h (adjustable)
Bolus rate, Prime rate (Purge)
999 ml/h (adjustable)
Infusion pressure minimal:
60 kPa / 450 mmHg / 0,6 Bar
Infusion pressure maximal:
120-150 kPa/ 900-1875 mmHg/
1.2-2.5 bar (depending on tubing)
Alarm pressure limit pressure sensor
0 - 999 mbar (mmHg optional)
Resolution pressure sensor
10 mbar (It is possible to choose
between “mbar” and “mmHg”).
Battery operation time (2.15Ah)
> 3h (up to 6 hours)
Charging time turned off/running
10 h
Supply voltage
110 - 230 VAC 50-60 Hz
External supply (optional)
12
– 15 V AC/DC
Current consumption
80 mA
Mains Fuse
500 mAT L /IEC127/III/SEV 1064
Electrical safety class
II (double insulation)
Protection against ingress of liquids and objects
IP32 (drip water proof, protected
against objects >= 2.5 mm)
Number of events saved:
1500, resist in memory after
switch off
Leakage current
< 40 µA
Radio interference
CE-Class B
Nurse call, potential-free
24 V/ 0,2 A
Classification
CF (cardiac floating)
Dimensions
(WxHxD) 245x90x180 mm Housing
ABS-plastic, UL listed
Weight
ca. 2,4 kg
Storage time
max. 3 months without charging
Atmospheric pressure
500-1000 hPa
Permitted temperature range (operation/storage):
15°C - 35°C / 0°C- 40°C
Permitted relative humidity:
20-90% (no vapour deposit)
Safety certification
IEC 60601-1
IEC 60601-2-24 / IEC 60601-1-2