35
GA 7000C-E-0313
© 2013 by arcomed ag, Switzerland
Setting up the wake up concentration:
While the pump is running enter the Menu, access the TCI mode and scroll until the awake
concentration menu is visible. Then by pressing one of the DATA keys the wake up
concentration can be set in DATA window. To confirm press key SELECT.
Figure 42: Setting the wake up concentration (wake up concentration: 1.00
g/ml)
3.3.8.3
Stopping the TCI Infusion
To stop or to interrupt the infusion press „Start/Stop“ key. Now the pump will display the time
until the patient should wake up (more precise: until the ce has fallen below the wake up
concentration). You can also set the TCI target concentration to zero to stop the TCI infusion.
This has the advantage that the pump will not issue a standby reminder alarm. However,
make sure this is really what you want. Crosscheck the displayed rate in the RATE window to
see if the pump is infusing or not. Note that depending on the TCI model, the rate might be
temporarily set to zero until the set target is reached and the pump will then continue to
infuse to maintain the target value.
Figure 43: pump not infusing cet = 0.00, time to wake up: 6:15 min, ce = 2.53
g/ml
For further information contact your
arcomed
dealer.