39
GA 7000C-E-0313
© 2013 by arcomed ag, Switzerland
panel. For safety reasons the touch screen is disabled indicating this with an arrow to the
panel.
Figure 49: Entering the bolus volume (1.0 ml), actual rate: 5.0 ml/h
Notes:
If no bolus is required, cancelling is possible by pressing ESC.
During delivery of the bolus, the bolus rate is displayed in the Rate window, the DATA
window displays the infused bolus volume. The pump can be stopped at any time by
pressing the „START/STOP“ key.
After the desired volume is delivered the pump continues with normal operation
mode, the delivered bolus volume is added to the volume already infused.
HINT:
The pump can be configured with a quick access (short cut) to the bolus function. This allows
speeding up and decreasing the number of key presses to enter the bolus menu. E.g if the
bolus is configured to be the OneClick short cut (indicated with a I at the bolus icon) the
following keys will trigger a bolus:
-
Manual Bolus: press I, press and hold II.
-
Automatic Bolus: press I, press I, enter bolus volume, press II.
Figure 50: Short cut to bolus