time. Then press and hold
“II” for as long the manual bolus shall be delivered. For safety
reasons the touch screen is disabled indicating this with an arrow to the panel. In the rate
window the bolus rate and in the DATA window the infused bolus volume is displayed. In the
LCD „
Bolus manual
“ is written. To stop the Bolus release “II”. The pump will continue in
normal operation mode.
Some drugs might be configured with a disabled bolus function for safety reasons. Contact
your pump administration to configure your pump to your needs.
3.4.5.2
Automatic Bolus
You will have also three possibilities to start an automatic bolus: using the Touch screen by
pressing MENU and Bolus/Prime icons, or using DATA keys and navigating to the
corresponding icon. Activate the bolus function by pressing “II”. To go to automatic bolus
press “I”. Thereafter the bolus speed is entered in the DATA window and confirmed by
pressing
SELECT or “II”.
Figure 47: Menu for choosing a bolus (actual rate: 100.0 ml/h)
After this the bolus speed is entered in the volume window and confirmed by pressing the
SELECT.
Figure 48: Setting the bolus speed (999 ml/h), actual rate: 5.0 ml/h
Next step is to enter the bolus volume in the DATA window. Now the bolus can be released
by pressing the
SELECT or “II”. Note that the bolus can only be started with the II key on the