51
GA 7000C-E-0313
© 2013 by arcomed ag, Switzerland
Typical trumpet curves at a rate of 42 ml/h with infusion set
“Volumed® -Set Standard”
(PVC):
T=2 h (full window)
2 Min
5 Min 11 Min 19 Min 31 Min
EP (Max)
3.44%
3.14%
3.00%
2.75%
2.59%
Rate = 42.71 ml/h
EP (Min)
0.37%
0.63%
0.67%
0.78%
0.90%
= 1.68 %
T=1.h (first hour)
2 Min
5 Min 11 Min 19 Min 31 Min
EP (Max)
3.44%
3.14%
3.00%
2.75%
2.59%
Rate = 42.91 ml/h
EP (Min)
0.64%
1.62%
1.63%
1.80%
1.90%
= 2.16 %
T=2.h (second hour)
2 Min
5 Min 11 Min 19 Min 31 Min
EP (Max)
2.08%
2.02%
1.86%
1.85%
1.95%
Rate = 42.51 ml/h
EP (Min)
0.37%
0.63%
0.67%
0.78%
0.90%
= 1.21 %
8
Chroma
Statement:
The Volumed® µVP7000 Chroma series is a medical electrical equipment and needs
special precautions regarding EMC and needs to be installed according to the EMC
information provided in this document.
Essential performance for Volumed® µVP7000 Chroma series are the following:
- infusing a certain volume of a medication at a certain rate.
Warning:
Portable and mobile RF communication equipment can affect medical electrical
equipment.
The Volumed® µVP7000 Chroma should not be used adjacent to or stacked with other
equipment. If adjacent or stacked use is necessary the Volumed® µVP7000 Chroma
series should be observed to verify normal operation in the configuration in which it will be
used.
Electrostatic discharges of more than 8kV can cause the pump or the display to switch off.
-10
-5
0
5
10
0
10
20
30
EP Max
EP Min
Total deviation
Trumpet curve for whole observation window
Ave
rag
e
d
ev
iatio
n
in
%
observation window in minutes
Содержание Volumed uVP7000 Premium