23
GA 7000C-E-0313
© 2013 by arcomed ag, Switzerland
4.
The rate is set in the corresponding unit in the DATA window. In the rate window the
rate is displayed in ml/h.
Figure 21: Setting the rate (1.00
g/kg/min, which equals 6.0 ml/h)
5.
After entering the last parameter you will get an overview screen to check again all
parameters.
Figure 21a: overview screen
6.
The rate in ml/h can always be checked by using the following formula:
1000
60
min
ml
mg
ion
concentrat
kg
weight
patient
kg
g
Rate
h
ml
Rate
7.
The
pump is started with the „START/STOP“ key.
NOTE:
When the pump is infusing, rate window will display the rate in ml/h. The LCD
displays the rate alternating in the corresponding unit and in ml/h.
Pressing SELECT twice will display additional information about the ongoing infusion
in the LCD.
To change the rate while the pump is running, press one of the volume or rate keys.
The rate window displays the speed in ml/h, the DATA window displays the rate in
g/kg/min. If the rate is changed, it is automatically converted to the other unit.