37
GA 7000C-E-0313
© 2013 by arcomed ag, Switzerland
3.4.4
Reset the Volume Infused
To reset the volume infused the key SELECT is pressed. Then press the DATA keys to scroll
to the Menu point „
Zero Volume
“ and press the SELECT key to select. In the LCD „
Zero
Volume infused?
“ is written. Press SELECT to confirm or ESC to cancel.
Figure 45: Menu to Zero the infused Volume
Figure 46: Confirmation to reset the volume infused
NOTE:
To reset the infused volume the pump must not be stopped. However, the volume
can only be reset with the II key on the panel for safety reasons.
3.4.5
Infusing a Bolus
While the pump is infusing a manual or automatic bolus can be delivered. There are several
ways to start the bolus. Be aware that for safety reasons the bolus can only be started with
the “II” key.
3.4.5.1
Manual Bolus:
Either the Menu icon and Bolus icon is pressed and the instruction on the screens are
followed or the DATA keys are used to navigate to the bolus icon. Activated the function by
pressing “II”. A third way to deliver a manual bolus, is pressing “II” and “I” together for a short