56
GA 7000C-E-0313
© 2013 by arcomed ag, Switzerland
Warning: After repairs, software updates and any replacement of parts, test runs must
be carried out in accordance with the
manufacturer’s protocols and documented with
the corresponding serial number of the devices.
The check list of the safety check is separately available at
arcomed ag
.
NOTE:
The original battery of the pump is
a NiMH Battery (9.6V, 2150mAh) and must be
checked. In case those other batteries than the ones supplied by
arcomed ag
are used, safe
operation of the pump cannot be guaranteed. Old batteries must be disposed in accordance
with national laws
arcomed
article number of the battery is 71160.
9.2
Repairs
The Volumed® µVP7000 Chroma infusion pump may only be serviced or repaired by
employees of
arcomed ag
or by people that were trained by
arcomed
employees. Please
contact your
arcomed
distributor for further information.
Warning: After a pump was at repair, please check the correct user specific settings
before using the pump on a patient.
Warning: Before opening the housing, make sure the pump is disconnected from
mains.