7
GA 7000C-E-0313
© 2013 by arcomed ag, Switzerland
1.4
Cleaning and Disinfection
Warning: The pump must be switched off and disconnected from the mains power
supply before cleaning and disinfecting.
The pump must be kept clean and dry. Remove any spillage immediately. The pump must
not be placed in an autoclave. Wipe over with a cloth, which has been damped slightly with
an alcohol-based disinfectant, to clean the unit. Take care when cleaning that no liquid
enters the inside of the pump case. Wait at least 30 seconds after disinfections before
switching the pump on. Use only disinfectants that are compliant with:
ABS, POM, stainless
steel, PVC, Aluminium, Silicone
(Please check with your supplier of disinfectant).
Disinfections, methylated Spirits and Alcohol with a concentration of more than 80 percent
are not allowed. Damages caused by misuse of cleaning agents are not covered by
warranty. It is not allowed to disinfect the pump by means of UV radiation! Please check the
user manual that comes with the disinfection agent to find out about the correct application
(residence time, dilution).
Contact between the cleaning agent and skin, eyes, mouth, or other body openings must be
avoided. In such a case the affected parts must be cleaned with a lot of clear water. A doctor
should be consulted in every case.
1.5
Key to Symbols
The pictograms and symbols shown on the reverse of the pump have the following meanings
or functions:
Device of Type CF (Cardiac Float)
Connection to automatic nurse
call station
Electrical safety class 2 (double
Insulation)
Declaration of conformity
IP32 protection against ingress of
liquids and objects
Manufacturer address
Consult user manual before using the
pump!
Connection for optional 12-15 V
AC/DC supply
Connection for Easy Clip
Dispose old pumps according to
the national guidelines. Old
batteries must be disposed off in
an environmentally friendly
manner or returned to the
manufacturer.
1.6
Power and Running Indicators
The window left to
the „ON/OFF“ key is used
to indicate the following states:
Pump infusing
Furthermore the LCD displays the
following:
If on: Pump on Mains operation
Else Pump on battery operation
Pump stopped
Pump infusing