19
GA 7000C-E-0313
© 2013 by arcomed ag, Switzerland
Figure 12: Choosing the drug Glucose 5% from the internal drug library
Furthermore it is possible to run the pump in a high precision mode (“
High Def Mode
”), which
allows entering rate and volume in steps of 0.01 ml/h (0.01 ml). The High Def Mode is
activated by selecting a drug defined as high definition drug.
Figure 13: High precision mode, Rate 6.25 ml/h, VTBI: 3.78ml, already infused: 0.59ml
3.3.5
Rate Calculation Using Drug Concentration and Patient Weight
The automatic rate calculation allows nursing staff to deliver drugs that are applied in
dependence of drug concentration and patient weight (optional). The drugs have to be
previously loaded into the pumps drug library.
To run the pump in this mode SELECT is pressed after the infusion set was loaded. Press
the DATA
keys to scroll to the menu point „
Drug Calculation
“.
Содержание Volumed uVP7000 Premium
Страница 1: ...Edition 03 2013 ...