6.15
MANUAL (AC) Mode Execution
In MANUAL mode of operation, the setup screen will look as shown here. The Manual mode
is indicated by the lack of a Step number after the M number in upper left corner of the LCD.
To enable the output and apply power to a unit under test, press the Test/Reset key and the
LED for the key will illuminate. The text “
Set
” on the set screen will turn to “
Dwell
”, in
addition, the soft keys will change to include “Meter”, “AUTO”, “Keypad” and “Trig.”.
In MANUAL Mode, the output will run continuously (infinite dwell time) until the Test/Reset
key is pressed again, or there is a failure condition. The Test/Reset LED will be on during this
time.
The following Soft keys are active during MANUAL mode output on state:
Meter
The “
Meter
” soft key allows the right hand side large font readout to be
toggle between all available measurements
AUTO / HIGH
The “
AUTO
” and “
HIGH
” soft keys appear in the same location and toggle
the voltage range between High and Low.
Note
that if the programmed
voltage is 150V or higher, the power source will remain in high voltage
range. If the programmed voltage is less than 150V, selecting AUTO will
result in the low voltage range being used allowing two times the amount of
current to be delivered to the unit under test.
Keypad
The “
Keypad
” soft key may be used to enter a different voltage or
frequency value directly using the Key pad. This means output voltage
and/or frequency can be changed without having to open the output (OFF).
You can also use the shuttle knob for this. Unless the Frequency is displayed
in large font on the right, the “
Keypad
” soft key and Shuttle both will adjust
the output voltage. If Frequency is displayed on the right, “Keypad” soft key
and shuttle will adjust the frequency.