SECTION 10: CALIBRATION
Entire Contents Copyright
2016 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS.
APS CFS100 Series Power Source Operation Manual
Page 142 of 146 PN 160951-10
10.4.7
Peak Current Calibration
Use the “
∧
,
∨
" soft keys to navigate to the A-Peak xx.xA parameter and press the
Select soft key.
Follow the prompt message provided on the display, and press the “
Test/Reset
” button to
move into the calibration screen for current. If you press the “
Exit
” soft key at this screen
you return to the calibration mode screen.
Enter the current reading from the ammeter with the numeric keypad. When the value has
been selected press the “
Enter
” soft key and you will be moved to the first calibration
parameter Volt 150.0V. The soft keys “
Volt+
” and “
Volt-
” are available, if needed, to adjust
the voltage output of the instrument to set a proper peak current value if a non-
recommended load is used. If you press the “
Esc
” soft key you will be returned to the
calibration mode screen
10.4.8
Exiting Calibration Mode
Once all calibrations are done, cycle power to the power source using the front panel
On/Off switch so the unit returns to normal operation mode.