5.10 Jitter Tolerance Setup Screen
5-83
5
R
em
ot
e C
on
trol
:SENSe:MEASure:BERCond:UNIT <type>
Parameter
<type>=<CHARACTER PROGRAM DATA>
RATE
Error rate
COUNt
Error count
ESTimate
Estimate
Function
Sets whether error rate, error count, or Estimate is used for pass/fail
judgment.
Example
To set the error rate for the Pass/Fail judgment:
> :SENSe:MEASure:BERCond:UNIT RATE
:SENSe:MEASure:BERCond:UNIT?
Response
<type>=<CHARACTER RESPONSE DATA>
RATE,COUN,EST
Function
Queries the setting of the Pass/Fail judgement
Example
> :SENSe:MEASure:BERCond:UNIT?
< RATE
:SENSe:MEASure:BERCond:THReshold <value>
Parameter
<value>=<DECIMAL NUMERIC PROGRAM DATA>
When Unit is set to Error Rate:
3 to 12
Error Rate E–3 to E–12/E–1 Step
When Unit is set to Error Count:
0 to 10000000
Number of Error Count 0 to 10000000/1Step
Function
Sets the evaluation threshold value.
Example
Sets the error rate threshold to 1E-9 when Unit is set to Error Rate.
> :SENSe:MEASure:BERCond:THReshold 9
Sets the error rate threshold to 100 when Unit is set to Error Count.
> :SENSe:MEASure:BERCond:THReshold 100
:SENSe:MEASure:BERCond:THReshold?
Response
<numeric>=<NR1 NUMERIC RESPONSE DATA>
When Unit is set to Error Rate:
3 to 12
Error Rate E–3 to E–12/E–1
When Unit is set to Error Count:
0 to 10000000
Error count 0 to 10000000
Function
Queries the evaluation threshold value of error rate
Example
> :SENSe:MEASure:BERCond:THReshold?
< 9
Содержание MX183000A
Страница 12: ...xii ...
Страница 16: ...IV ...
Страница 20: ...Chapter 1 Outline 1 4 Figure 1 1 3 Jitter Tolerance Graph Screen ...
Страница 54: ...Chapter 3 Connecting Equipment 3 16 ...
Страница 136: ...Chapter 5 Remote Control 5 30 Figure 5 8 2 Sequence Screen BER Measurement 10 8 9 11 ...
Страница 206: ...Chapter 5 Remote Control 5 100 ...
Страница 218: ...Appendix A Specifications A 12 ...
Страница 226: ...Appendix B Default Settings B 8 ...