5.2 Remote Control Procedure
5-9
5
R
em
ot
e C
on
trol
This is an example of the procedure for sending a Link Sequence using a
USB device.
1. Start the USB Link Sequence application.
:SYSTem:MEASure:SELect USB
2. Connect using steps 6 to 9 in the jitter tolerance test example
procedure.
3. Set the test target specification to GEN 1.
:DISPlay:MEASure:CHANge SEQuence
:LTRaining:SEQuence:SPECification GEN1
4. Set the test pattern to Compliance Pattern CP0.
:SOURce:PATTern:TYPE COMPliance
:LTRaining:SEQuence:TEST:PATTern 0
5. Send a sequence to loopback the DUT. *
:LTRaining:SEQuence:STARt
*: When this command is sent, the DUT should be connected to
the USB test fixture.
6. During sequence transmission, the only remote control operations
possible are measurement stop and obtain measurement status.
:LTRaining:SEQuence:STATe?
:LTRaining:SEQuence:STOP
7. Close the application.
:LTRaining:SEQuence:STOP
:SYSTem:MEASure:SELect NONE