Chapter 5 Remote Control
5-68
:SENSe:MEASure:BER:STOP
Parameter
None
Function
Starts the BER measurement.
Example
>:SENSe:MEASure:BER:STOP
:SENSe:MEASure:BER:STATe?
Response
<numeric>=<NR1 NUMERIC RESPONSE DATA>
0
Stop
1
Measurement in progress
Function
Queries the BER measurement status.
Example
>:SENSe:MEASure:BER:STATe?
< 1
:DISPlay:RESult:BER <boolean>
Parameter
<boolean>=<BOOLEAN PROGRAM DATA>
OFF or 0
Hides the BER measurement setting window.
ON or 1
Displays the BER measurement setting window.
Function
Displays or hides the BER measurement setting window.
Example
To display the BER measurement setting window
>:DISPlay:RESult:BER ON
:DISPlay:RESult:BER?
Response
<numeric>=<NR1 NUMERIC RESPONSE DATA>
0
Hides the BER measurement setting window.
1
Displays the BER measurement setting window.
Function
Queries the BER measurement setting window display status.
Example
>:DISPlay:RESult:BER?
< 1
:SENSe:MEASure:BER:ECTHreshold <numeric>
Parameter
<numeric>=<DECIMAL NUMERIC PROGRAM DATA>
0 to 1000
0 to 1000/1step
Function
Sets an evaluation threshold of the BER measurement.
When the number of bit errors exceeds the evaluation threshold, the BER
measurement is judged as Fail.
Example
To set the evaluation threshold of the BER measurement to 1:
>:SENS:MEAS:BER:ECTHreshold 1
Содержание MX183000A
Страница 12: ...xii ...
Страница 16: ...IV ...
Страница 20: ...Chapter 1 Outline 1 4 Figure 1 1 3 Jitter Tolerance Graph Screen ...
Страница 54: ...Chapter 3 Connecting Equipment 3 16 ...
Страница 136: ...Chapter 5 Remote Control 5 30 Figure 5 8 2 Sequence Screen BER Measurement 10 8 9 11 ...
Страница 206: ...Chapter 5 Remote Control 5 100 ...
Страница 218: ...Appendix A Specifications A 12 ...
Страница 226: ...Appendix B Default Settings B 8 ...