Rev. 5
43
Operation
Shift Key Commands : Field Units
3.7.7 Field
Units
SHIFT-key
Use of the
FIELD UNITS
SHIFT-key provides a shortcut to the picklist
menu
1
for defining whether the field is specified and displayed in units of
kilogauss (kG) or tesla (T). The selected option also applies to remote
interface commands. The default setting is kilogauss.
3.7.8
Persistent Switch Heater Current SHIFT-key
Use of the
P. SWITCH HTR. CUR.
SHIFT-key provides a shortcut to the
menu
2
for setting persistent switch heater current. The value can be set to
between 0.0 and 125.0 mA. The default value is 10.0 mA unless preset by
AMI to match a specific superconducting magnet.
3.7.9 Stability
SHIFT-key
Use of the
STAB.
SHIFT-key provides a shortcut to the menu for defining
the Model 430 stability setting. The stability setting is specified in percent
and controls the transient response and stability of the system. The valid
input range is from 0.0 to 100.0%. The default value is 0.0% unless preset
by AMI to match a specific superconducting magnet.
See section 3.10.2.1 on page 52 for details of how to determine the stability
setting to use.
3.7.10 Vs <> Vm SHIFT-key
The
Vs <> Vm
SHIFT-key is used to toggle the voltage display between
display of the voltage across the magnet (Vm) and the power supply output
voltage (Vs). See section 3.2.2 on page 28 for details.
3.7.11 Volt
Meter
SHIFT-key
The
VOLT METER
SHIFT-key is used to toggle the main display between
display of a voltmeter indicating magnet voltage (Vm) or supply voltage
(Vs), and display of ramp mode and persistent switch heater state. See
section 3.2.4 on page 29 for details.
1. Refer to section 3.10.3.4 on page 64.
2. Refer to section 3.10.2.8 on page 59.
+50.00 A - Field Units
+0.50 Vs
Kilogauss
Tesla
+50.00 A — PSwitch Current (mA)
+0.50 Vs
10.0
Содержание 05100PS-430-601
Страница 2: ......
Страница 10: ...x Rev 5 List of Figures ...
Страница 12: ...xii Rev 5 List of Tables ...
Страница 18: ...xviii Rev 5 Foreword Safety Summary ...
Страница 30: ...12 Rev 5 Introduction Operating Characteristics ...
Страница 42: ...24 Rev 3 Installation Power Up Procedure ...
Страница 114: ...96 Rev 5 Operation Summary of Operational Limits ...
Страница 119: ...Rev 5 101 Remote Interface Reference SCPI Command Summary LOCK ABsorber LOCK BRIGHTness LOCK NETsetup ...
Страница 156: ...138 Rev 5 Remote Interface Reference Error Messages ...
Страница 168: ...150 Rev 5 Service Return Authorization ...
Страница 190: ...172 Rev 5 Appendix Power Supply Details Figure A 3 Model 08150PS Dimensions Top and Side Views ...
Страница 220: ...202 Rev 5 Appendix Persistent Switch Operation Flowchart Figure A 17 Persistent Switch Operation Flowchart Page 3 ...
Страница 226: ...208 Rev 5 Index ...