52
Rev. 5
Operation
Setup Submenu : Load
3.10.1.1.1.4 Max Output Current
The maximum output current is specified in amperes (A) and
reflects the maximum output current capacity of a connected power
supply. The valid range is 0.001 to +2000.000 A
1
, and can be set by
using either the numeric keypad per section 3.3 on page 29 or the
fine adjust knob (section 3.4 on page 31).
3.10.1.1.1.5 V-V Mode Input Range
The voltage-to-voltage mode input range defines the remote
programming voltage input range required by the connected power
supply. The remote programming voltage is the output signal
provided by the Model 430 Programmer as an input to the
connected power supply.
This submenu item provides a picklist of six preset selections and
does not allow numeric entry of a range. The picklist values are
shown in Table 3-3 below.
3.10.2 Load
Submenu
When the Load submenu is selected, several parameters associated with
the superconducting magnet load can be viewed and/or changed.
3.10.2.1
Stability Setting
The stability setting is specified in percent and controls the
transient response and stability of the system. The value can be set
Table 3-3.
V-V Mode Input Range Picklist Values
+0.000 to -5.000
+0.000 to +8.000
+0.000 to +5.000
-5.000 to +5.000
+0.000 to +10.000
-10.000 to +10.000
+0.00 A — Max Output Current (A)
+0.00 Vs
+100.000
+0.00 A — V-V Mode Input Range (V)
+0.00 Vs
-10.000 to +10.000
+0.00 A — Stability Setting (%)
+0.50 Vs
0.0
Содержание 05100PS-430-601
Страница 2: ......
Страница 10: ...x Rev 5 List of Figures ...
Страница 12: ...xii Rev 5 List of Tables ...
Страница 18: ...xviii Rev 5 Foreword Safety Summary ...
Страница 30: ...12 Rev 5 Introduction Operating Characteristics ...
Страница 42: ...24 Rev 3 Installation Power Up Procedure ...
Страница 114: ...96 Rev 5 Operation Summary of Operational Limits ...
Страница 119: ...Rev 5 101 Remote Interface Reference SCPI Command Summary LOCK ABsorber LOCK BRIGHTness LOCK NETsetup ...
Страница 156: ...138 Rev 5 Remote Interface Reference Error Messages ...
Страница 168: ...150 Rev 5 Service Return Authorization ...
Страница 190: ...172 Rev 5 Appendix Power Supply Details Figure A 3 Model 08150PS Dimensions Top and Side Views ...
Страница 220: ...202 Rev 5 Appendix Persistent Switch Operation Flowchart Figure A 17 Persistent Switch Operation Flowchart Page 3 ...
Страница 226: ...208 Rev 5 Index ...