14
Rev.
3
Installation
Power Requirements
2.3 Power Requirements
Warning
The system operates on 50/60 Hz power and may be configured for
100-115 VAC or 200-230 VAC. The power requirement for each
system component is marked on the rear panel of the unit adjacent
to the power entry connector. Be sure the power supply system is
configured for the proper power source prior to plugging in the line
cords. Do not fail to connect the input ground terminal securely to
an external earth ground.
Ensure the front panel power switches are in the OFF (
O
) position. Verify
that the power supply components are configured for the proper operating
voltage by referring to the equipment rear panels. If the operating voltage
is correct, plug the line cords into power entry connectors, and into the
appropriate power receptacles.
2.3.1 Changing the Model 430 Programmer Operating Voltage
Warning
The following procedure is to be performed only when the Model 430
Programmer is completely de-energized by removing the power-cord
from the power receptacle. Failure to do so could result in personnel
coming in contact with high voltages capable of producing life-
threatening electrical shock.
Note
The voltage selector switch is labeled “115” for nominal line voltages
from 100 to 115 VAC. The switch is labeled “230” for nominal line
voltages of 200 to 230 VAC.
If the Model 430 Programmer operating voltage must be changed, ensure
the instrument is de-energized by disconnecting the power cord from the
power source. Remove the Model 430 Programmer cover by removing the
four screws on both sides of the cover and the four screws from the corners
of the cover on the back panel; slide the voltage selector switch on the main
printed circuit board to the proper voltage. Replace the Model 430
Programmer cover.
2.4 Collecting Necessary Information
In order to properly configure the Model 430 Programmer, specific system
information is required. Such parameters as the magnet electrical
properties, type of power supply, persistent switch heating current
requirements, and voltage and current constraints of the magnet are
entered into the Model 430 Programmer once and nonvolatile memory will
Содержание 05100PS-430-601
Страница 2: ......
Страница 10: ...x Rev 5 List of Figures ...
Страница 12: ...xii Rev 5 List of Tables ...
Страница 18: ...xviii Rev 5 Foreword Safety Summary ...
Страница 30: ...12 Rev 5 Introduction Operating Characteristics ...
Страница 42: ...24 Rev 3 Installation Power Up Procedure ...
Страница 114: ...96 Rev 5 Operation Summary of Operational Limits ...
Страница 119: ...Rev 5 101 Remote Interface Reference SCPI Command Summary LOCK ABsorber LOCK BRIGHTness LOCK NETsetup ...
Страница 156: ...138 Rev 5 Remote Interface Reference Error Messages ...
Страница 168: ...150 Rev 5 Service Return Authorization ...
Страница 190: ...172 Rev 5 Appendix Power Supply Details Figure A 3 Model 08150PS Dimensions Top and Side Views ...
Страница 220: ...202 Rev 5 Appendix Persistent Switch Operation Flowchart Figure A 17 Persistent Switch Operation Flowchart Page 3 ...
Страница 226: ...208 Rev 5 Index ...