114
Rev. 5
Remote Interface Reference
SETUP Configuration Commands and Queries
•
*CLS
Clears the Standard Event register and the error buffer.
•
*ESR?
Returns a decimal sum which corresponds to the binary-weighted sum of
the contents of the Standard Event register.
•
*ESE <
enable_value
>
Enables bits in the Standard Event register to be reported in the
“Standard Event” bit (bit 5) of the Status Byte register. To enable bits, you
must write a decimal
<
enable_value
>
which corresponds to the binary-
weighted sum of the bits you wish to enable. Refer to Table 4-2 on page 108
for more information. For example, to enable
all
categories of error
messages to be reported in bit 5 of the Status Byte register, send:
*ESE 60;
•
*ESE?
The
*ESE?
query returns a decimal sum which corresponds to the binary-
weighted sum of the bits enabled by the last
*ESE
command.
•
*PSC {0|1}
Power-On Status Clear
. If
*PSC 1
is in effect, the Standard Event enable
register and the Status Byte enable register are cleared at power on. If
*PSC 0
is in effect, the enable registers are not cleared at power on. The
default setting is “1”.
•
*PSC?
Returns the
Power-On Status Clear
setting currently in effect. A value of
“0” indicates the enable registers are not cleared at power on; a value of “1”
indicates the enable registers are cleared at power on.
•
*OPC
Sets the “Operation Complete” bit (bit 0) of the Standard Event register
when executed. See page 108 for a complete discussion.
•
*OPC?
Returns “1” to the requesting interface when executed. See page 109 for
more information.
4.5.3
SETUP Configuration Commands and Queries
The SETUP Configuration Commands and Queries provide read/write
access to the setup functions available within the Supply (see section
(see section 3.10.2), and Misc
(see section 3.10.3) submenus.
Содержание 05100PS-430-601
Страница 2: ......
Страница 10: ...x Rev 5 List of Figures ...
Страница 12: ...xii Rev 5 List of Tables ...
Страница 18: ...xviii Rev 5 Foreword Safety Summary ...
Страница 30: ...12 Rev 5 Introduction Operating Characteristics ...
Страница 42: ...24 Rev 3 Installation Power Up Procedure ...
Страница 114: ...96 Rev 5 Operation Summary of Operational Limits ...
Страница 119: ...Rev 5 101 Remote Interface Reference SCPI Command Summary LOCK ABsorber LOCK BRIGHTness LOCK NETsetup ...
Страница 156: ...138 Rev 5 Remote Interface Reference Error Messages ...
Страница 168: ...150 Rev 5 Service Return Authorization ...
Страница 190: ...172 Rev 5 Appendix Power Supply Details Figure A 3 Model 08150PS Dimensions Top and Side Views ...
Страница 220: ...202 Rev 5 Appendix Persistent Switch Operation Flowchart Figure A 17 Persistent Switch Operation Flowchart Page 3 ...
Страница 226: ...208 Rev 5 Index ...