Rev. 3
15
Installation
Bipolar High-Stability Supply
retain the data even after power is removed from the instrument. An
example of the data to be entered and how it is entered is described in
If the Model 430 Programmer was purchased as part of a magnet system,
essential data will have already been entered at the AMI factory and a
configuration sheet will have been provided detailing the settings.
2.5 System Interconnects
If the Model 430 Programmer was purchased as part of a magnet system,
all applicable system components and wiring harnesses will have been
shipped with the system.
Multiple power supplies should be connected to the same AC input power
source. The protection is configured so that a fault will shut down all the
interconnected power supplies.
The diagrams that follow will assist in system equipment setup.
Caution
The wiring between the power supply and the magnet current leads
must be of sufficient size to carry the full rated current of the power
supply. Typically, for short runs (less than 25 ft, or 7.6 m), 2 AWG
wire is sufficient for 125 A current, and 2/0 AWG wire is best for
250 A current.
Note that an AMI Model 13x Liquid Helium Level Instrument is shown as
a possible component of each integrated system. The main
instrumentation cable connecting the magnet support stand to one of the
Model 430 Programmer
MAGNET STATION
connectors contains all the
instrumentation and control connections needed to control and monitor
the magnet. The signals in this cable which are required to monitor LHe
level and temperatures are also presented at the LHe Level / Temp
Connectors. Refer to the
Appendix
for pin-outs of these and other
connectors.
2.5.1 High-Stability Bipolar Supply
Current stability of a system can be increased by an order of magnitude
through application of the zero flux method of current sensing. For the
bipolar (dual-quadrant) mode high-stability magnet power supply, the
system consists of the Model 430 Programmer with zero flux current
sensing, unipolar 08150PS Power Supply, Model 601 Energy Absorber,
and associated interconnection cabling.
Содержание 05100PS-430-601
Страница 2: ......
Страница 10: ...x Rev 5 List of Figures ...
Страница 12: ...xii Rev 5 List of Tables ...
Страница 18: ...xviii Rev 5 Foreword Safety Summary ...
Страница 30: ...12 Rev 5 Introduction Operating Characteristics ...
Страница 42: ...24 Rev 3 Installation Power Up Procedure ...
Страница 114: ...96 Rev 5 Operation Summary of Operational Limits ...
Страница 119: ...Rev 5 101 Remote Interface Reference SCPI Command Summary LOCK ABsorber LOCK BRIGHTness LOCK NETsetup ...
Страница 156: ...138 Rev 5 Remote Interface Reference Error Messages ...
Страница 168: ...150 Rev 5 Service Return Authorization ...
Страница 190: ...172 Rev 5 Appendix Power Supply Details Figure A 3 Model 08150PS Dimensions Top and Side Views ...
Страница 220: ...202 Rev 5 Appendix Persistent Switch Operation Flowchart Figure A 17 Persistent Switch Operation Flowchart Page 3 ...
Страница 226: ...208 Rev 5 Index ...