Rev. 5
9
Introduction
Model 601 Front Panel
1.6 System Specifications @ 25
°
C
Magnet Current Control
Range:
0 to +100 A
Programming Accuracy:
5 mA
Stability:
1 mA after 10 min. at desired current
Minimum Ramp Rate:
100
μ
A/min
Maximum Ramp Rate:
10 A/sec
Output Voltage
Range:
0 to
±
5 Vdc
Measurement Resolution:
10 mV
Load Inductance
Range:
0.5 to 100 H
Primary Power Requirements
Range:
100 - 115 or 200 - 230 Vac
±
10%
50 / 60 Hz, 1500 VA
Physical
Dimensions
a
:
a. H = height; W = width; D = depth
12.5” H x 21” W x 24.5” D
(318 mm H x 533 mm W x 622 mm D)
Approximate Weight:
80 lbm (35 kg)
Terminal Torque Limit:
48 lbf-in (5.4 N-m)
Environmental Limits
Ambient Temperature:
0 °C to 40 °C (32 °F to 104 °F)
Relative Humidity:
0 to 95%; non-condensing
Содержание 05100PS-430-601
Страница 2: ......
Страница 10: ...x Rev 5 List of Figures ...
Страница 12: ...xii Rev 5 List of Tables ...
Страница 18: ...xviii Rev 5 Foreword Safety Summary ...
Страница 30: ...12 Rev 5 Introduction Operating Characteristics ...
Страница 42: ...24 Rev 3 Installation Power Up Procedure ...
Страница 114: ...96 Rev 5 Operation Summary of Operational Limits ...
Страница 119: ...Rev 5 101 Remote Interface Reference SCPI Command Summary LOCK ABsorber LOCK BRIGHTness LOCK NETsetup ...
Страница 156: ...138 Rev 5 Remote Interface Reference Error Messages ...
Страница 168: ...150 Rev 5 Service Return Authorization ...
Страница 190: ...172 Rev 5 Appendix Power Supply Details Figure A 3 Model 08150PS Dimensions Top and Side Views ...
Страница 220: ...202 Rev 5 Appendix Persistent Switch Operation Flowchart Figure A 17 Persistent Switch Operation Flowchart Page 3 ...
Страница 226: ...208 Rev 5 Index ...