Rev. 5
33
Operation
Single-key Commands
selector will move to the last picklist value. An example of a picklist entry
in progress (picklist entry active) is illustrated below:
When the item selector is pointing at the desired picklist value, press the
ENTER
key to accept the picklist value. Values are
not
applied to the
operation of the Model 430 Programmer until the
ENTER
key is pressed
and the asterisk disappears from the display.
If the
ESC
key is pressed while picklist entry is active, the setting will
revert to the previous value and picklist entry will be made inactive. Note
that if the
ESC
key is pressed when numeric entry is
not
active, the
current submenu will be exited and the next higher level submenu will be
entered (if it exists).
3.6
Single-key Commands / Menu
All ramping controls (
PERSIST. SWITCH CONTROL
,
TARGET FIELD
SETPOINT
,
RAMP / PAUSE
and
RAMP TO ZERO
) are accessed with a
single keystroke. See section 3.12 on page 79 for details of ramping
controls. Below is a brief summary of the function of each of these keys.
3.6.1
Persistent Switch Control Key
Pressing the
PERSIST. SWITCH CONTROL
key toggles the Model 430
Programmer persistent switch heater control function.
If the persistent switch heater is energized and this key is pressed, the
persistent switch heater is de-energized. The power supply is then
maintained for the time set by the PSwitch Cooled Time variable before
being ramped down to zero at the rate set by the PSw P/S Ramp Rate
variable (default rate is 10 A/sec).
If the persistent switch heater is de-energized and this key is pressed, the
power supply is ramped to the current present in the magnet when the
+50.00 A -
Field Units*
+0.50 Vs
Kilogauss
Tesla
Figure 3-4.
Single Input Keys
Содержание 05100PS-430-601
Страница 2: ......
Страница 10: ...x Rev 5 List of Figures ...
Страница 12: ...xii Rev 5 List of Tables ...
Страница 18: ...xviii Rev 5 Foreword Safety Summary ...
Страница 30: ...12 Rev 5 Introduction Operating Characteristics ...
Страница 42: ...24 Rev 3 Installation Power Up Procedure ...
Страница 114: ...96 Rev 5 Operation Summary of Operational Limits ...
Страница 119: ...Rev 5 101 Remote Interface Reference SCPI Command Summary LOCK ABsorber LOCK BRIGHTness LOCK NETsetup ...
Страница 156: ...138 Rev 5 Remote Interface Reference Error Messages ...
Страница 168: ...150 Rev 5 Service Return Authorization ...
Страница 190: ...172 Rev 5 Appendix Power Supply Details Figure A 3 Model 08150PS Dimensions Top and Side Views ...
Страница 220: ...202 Rev 5 Appendix Persistent Switch Operation Flowchart Figure A 17 Persistent Switch Operation Flowchart Page 3 ...
Страница 226: ...208 Rev 5 Index ...