Introduction
Model 601 Front Panel
1.5 Model 601 Energy Absorber Front Panel Layout
The Fault LED is the only device on the Model 601 front panel. If the Fault
LED is
not
energized, the Model 601 is operating correctly. If the Fault
LED is energized, then one or more of the internal energy absorbing
elements has malfunctioned or power has been lost to the rear-panel
power connector. An audible alarm will also sound when the Fault LED is
energized.
Caution
If the system is in operation when an energy absorber fault occurs, a
safe magnet system state (typically zero current or a cooled
persistent switch in a connected magnet) should immediately be
attained. Do not continue to operate the unit, and refer to the
“Troubleshooting Hints” on page 139 for further direction.
Figure 1-3.
Model 601 Front Panel Layout
Содержание 05100PS-430-601
Страница 2: ......
Страница 10: ...x Rev 5 List of Figures ...
Страница 12: ...xii Rev 5 List of Tables ...
Страница 18: ...xviii Rev 5 Foreword Safety Summary ...
Страница 30: ...12 Rev 5 Introduction Operating Characteristics ...
Страница 42: ...24 Rev 3 Installation Power Up Procedure ...
Страница 114: ...96 Rev 5 Operation Summary of Operational Limits ...
Страница 119: ...Rev 5 101 Remote Interface Reference SCPI Command Summary LOCK ABsorber LOCK BRIGHTness LOCK NETsetup ...
Страница 156: ...138 Rev 5 Remote Interface Reference Error Messages ...
Страница 168: ...150 Rev 5 Service Return Authorization ...
Страница 190: ...172 Rev 5 Appendix Power Supply Details Figure A 3 Model 08150PS Dimensions Top and Side Views ...
Страница 220: ...202 Rev 5 Appendix Persistent Switch Operation Flowchart Figure A 17 Persistent Switch Operation Flowchart Page 3 ...
Страница 226: ...208 Rev 5 Index ...