Remote Interface Reference
SETUP Configuration Commands and Queries
•
ABsorber?
Returns “0” indicating that an energy absorber is not present in the
system, or “1” indicating that an energy absorber is present.
•
CONFigure:RAMP:RATE:UNITS {0|1}
Sets the preferred ramp rate time units. Sending “0” selects seconds. A “1”
selects minutes. “0” is the default value. The selected units are applied to
both the Model 430 Programmer display and the appropriate remote
commands.
•
RAMP:RATE:UNITS?
Returns “0” for ramp rates displayed/specified in terms of seconds, or “1”
for minutes.
•
CONFigure:FIELD:UNITS {0|1}
Sets the preferred field units. Sending “0” selects kilogauss. A “1” selects
tesla. “0” is the default value. The selected field units are applied to both
the Model 430 Programmer display and the applicable remote commands.
•
FIELD:UNITS?
Returns “0” for field values displayed/specified in terms of kilogauss, or “1”
for tesla.
•
CONFigure:IPNAME <
system name
>
Sets the system name (also known as
host name
or
computer name
), the
name
by which the Model 430 Programmer is identified on a network.
Note
If the system name value is changed, the Model 430 Programmer
power must be cycled off for at least 5 seconds and then back on to
complete the change. The new value will be used internally
immediately (even before cycling power off and back on), but the
IPNAME? query will return the previous system name until the
Model 430 Programmer is restarted.
•
IPNAME?
Returns the system name (also known as
host name
or
computer name
).
4.5.4
Protection Commands and Queries
The Protection Commands and Queries provide read/write access to all
protection settings which can be accessed under the Settings Protection
submenu. See section 3.10.3.6 for more information regarding the settings
protection features of the Model 430 Programmer.
Содержание 05100PS-430-601
Страница 2: ......
Страница 10: ...x Rev 5 List of Figures ...
Страница 12: ...xii Rev 5 List of Tables ...
Страница 18: ...xviii Rev 5 Foreword Safety Summary ...
Страница 30: ...12 Rev 5 Introduction Operating Characteristics ...
Страница 42: ...24 Rev 3 Installation Power Up Procedure ...
Страница 114: ...96 Rev 5 Operation Summary of Operational Limits ...
Страница 119: ...Rev 5 101 Remote Interface Reference SCPI Command Summary LOCK ABsorber LOCK BRIGHTness LOCK NETsetup ...
Страница 156: ...138 Rev 5 Remote Interface Reference Error Messages ...
Страница 168: ...150 Rev 5 Service Return Authorization ...
Страница 190: ...172 Rev 5 Appendix Power Supply Details Figure A 3 Model 08150PS Dimensions Top and Side Views ...
Страница 220: ...202 Rev 5 Appendix Persistent Switch Operation Flowchart Figure A 17 Persistent Switch Operation Flowchart Page 3 ...
Страница 226: ...208 Rev 5 Index ...