84
Rev. 5
Operation
Persistent Switch Control : Entering Persistent Mode
Note
The magnet voltage (Vm) is monitored during the power supply
ramp to zero. If the magnet voltage exceeds 0.5 V during this ramp,
the ramp is paused and the Model 430 Programmer beeps to
indicate the persistent switch did not transition to the
superconducting state properly. If this error occurs, the Model 430
will ramp the current back to the value when the persistent switch
was cooled,
and then heat the switch:
After the persistent switch heated time has elapsed, the display
will indicate the persistent mode transition malfunction:
After pressing
ENTER
, the Model 430 Programmer will revert to
the default field/current display.
5. After the power supply is finished ramping to zero, the following
screen will be displayed:
+45.39 A
Mode: Power Supply Ramping
+3.44 Vs
to match magnet field
+50.00 A
Mode: Heating Switch
+3.50 Vs
***** PSW Lock Error *****
+50.00 A -
Mode: Paused w/PSW error
+3.50 Vs
Press ENTER to continue
+50.00 A
P
Mode: Paused
+3.50 Vs
PSwitch Heater: ON
+0.00 A -
Magnet in Persistent Mode
+0.00 Vm
Press PER.SW.CTRL. to exit
Содержание 05100PS-430-601
Страница 2: ......
Страница 10: ...x Rev 5 List of Figures ...
Страница 12: ...xii Rev 5 List of Tables ...
Страница 18: ...xviii Rev 5 Foreword Safety Summary ...
Страница 30: ...12 Rev 5 Introduction Operating Characteristics ...
Страница 42: ...24 Rev 3 Installation Power Up Procedure ...
Страница 114: ...96 Rev 5 Operation Summary of Operational Limits ...
Страница 119: ...Rev 5 101 Remote Interface Reference SCPI Command Summary LOCK ABsorber LOCK BRIGHTness LOCK NETsetup ...
Страница 156: ...138 Rev 5 Remote Interface Reference Error Messages ...
Страница 168: ...150 Rev 5 Service Return Authorization ...
Страница 190: ...172 Rev 5 Appendix Power Supply Details Figure A 3 Model 08150PS Dimensions Top and Side Views ...
Страница 220: ...202 Rev 5 Appendix Persistent Switch Operation Flowchart Figure A 17 Persistent Switch Operation Flowchart Page 3 ...
Страница 226: ...208 Rev 5 Index ...