Ramp Configuration Commands and Queries
•
FIELD:MAGnet?
Returns the calculated field in kilogauss or tesla, per the selected field
units. This query requires that a coil constant be defined; otherwise, an
error is generated. The field is calculated by multiplying the measured
magnet current by the coil constant. If the magnet is in persistent mode,
the command returns the field that was present when persistent mode was
entered.
•
INDuctance?
Returns the measured magnet inductance in henries. Note that the
magnet must be ramping when this command is executed. Refer to section
•
CONFigure:RAMPDown:ENABle {0|1}
Enables the external rampdown function. “1” enables while “0” disables.
“0” is the default value.
•
RAMPDown:ENABle?
Queries whether the external rampdown function is enabled. Returns “1”
for enabled while “0” for disabled. “0” is the default value.
•
CONFigure:RAMPDown:RATE:SEGments <
# segments
>
Sets the number of external rampdown segments.
•
RAMPDown:RATE:SEGments?
Returns the number of external rampdown segments.
•
CONFigure:RAMPDown:RATE:CURRent <
segment
>,<
rate (A/s, A/
min)
>, <
upper bound (A)
>
Sets the external rampdown rate for the specified segment (values of 1
through the defined number of rampdown segments are valid) in units of
A/sec or A/min (per the selected rampdown rate units), and defines the
current upper bound for that segment in amperes.
•
RAMPDown:RATE:CURRent: <
segment
>?
Returns the external rampdown rate setting for the specified segment
(values of 1 through the defined number of rampdown segments are valid)
in units of A/sec or A/min (per the selected rampdown rate units) and the
current upper bound for that range in amperes. The two return values are
separated by a comma. For example:
RAMPDown:RATE:CURRENT:1?
0.1000,50.0000
Содержание 05100PS-430-601
Страница 2: ......
Страница 10: ...x Rev 5 List of Figures ...
Страница 12: ...xii Rev 5 List of Tables ...
Страница 18: ...xviii Rev 5 Foreword Safety Summary ...
Страница 30: ...12 Rev 5 Introduction Operating Characteristics ...
Страница 42: ...24 Rev 3 Installation Power Up Procedure ...
Страница 114: ...96 Rev 5 Operation Summary of Operational Limits ...
Страница 119: ...Rev 5 101 Remote Interface Reference SCPI Command Summary LOCK ABsorber LOCK BRIGHTness LOCK NETsetup ...
Страница 156: ...138 Rev 5 Remote Interface Reference Error Messages ...
Страница 168: ...150 Rev 5 Service Return Authorization ...
Страница 190: ...172 Rev 5 Appendix Power Supply Details Figure A 3 Model 08150PS Dimensions Top and Side Views ...
Страница 220: ...202 Rev 5 Appendix Persistent Switch Operation Flowchart Figure A 17 Persistent Switch Operation Flowchart Page 3 ...
Страница 226: ...208 Rev 5 Index ...