MU-100A
10. Setting of Measurement Mode and Schedule
10-15
4
Set No.2 “Weld-Time” to the time longer than the set time for the welding power
supply.
5
Set the monitoring level according to your use conditions.
When No.2 “Weld-Time” is shorter than the setting of No.28 “Force Sampling”
or “Force Average”, the setting is not done.
Monitoring of the force during welding
No.12 “After Weld Force LO” (after-welding force lower limit)
No.13 “After Weld Force UP” (after-welding force upper limit)
6
Set the level output according to your use conditions.
Use of the force level output
No.17 “Force Level1”
No.18 “Force Level2”
No.19 “Force Level3”
7
Set the weld start output force value to your use conditions.
Setting of the weld start output force value
No.21 “Tangential Force”
c. Measurement
1
Press MONI key to display the normal screen.
2
Set the schedule by selecting schedule switching signals “SCH1”, “SCH2”,
“SCH4”, “SCH8”, “SCH16”, “SCH32”, and “SCH64” or via the panel.
3
Apply the force to the load cell with the ready state. When the force value
reached the setting of No.21 “Tangential Force”, the sequence moves to
Weld-Time and the measurement is done.
“ST-FORCE1” or “ST-FORCE2” (second axis) is not used. Also, two axes do
not start simultaneously.