B 120
Alcatel Vacuum Technology France - ATP User’s Manual
Detailed description of RS commands
GB 00420 - Edition 03 - March 00
CYC
Syntax
Result
CKS
Syntax
Result
Enables or disables reply strings checksum
#adr
CKS
ON<CR>[<LF>]
Enables ascii character checksum at the end of
a reply string
or
#adr
CKS
OFF<CR>[<LF>]
Disables ascii character checksum at the end of
a reply string
#adr,ok,
S
for CKSON
#adr,ok
for CKSOFF
This feature allows the user to test if there is any transmit error
with a reply string.
S
is a character whose ascii value is the checksum, on 7 bits,
of all the character ascii values from the beginning of the reply
string to the character before
S
. The 8th bit of
S
(MSB, Most
Significant Bit) is always 1.
Starts the specified running-in cycle
#adr
CYC
1<CR>[<LF>] to start running-in program 1,
or
#adr
CYC
2<CR>[<LF>] to start running-in program 2
#adr,ok
Running-in program 1 should be executed after a pump maintenance
operation (change of bearings).
At the end of the program, the pump maintenance parameters are
updated and the «maintenance requested» alert can be cleared.
Program 2 is used after regreasing (ATP series only), or after
prolonged storage (ATH 20/40 only).