5/10
B 120
Alcatel Vacuum Technology France - ATP User’s Manual
Detailed description of RS commands
GB 00420 - Edition 03 - March 00
LEV
Syntax
Result
Syntax
Result
See also:
LNG, SEP, SHT
Returns the state of the parameters defined by SET
#adr
LEV
<CR>[<LF>]
#adr,nnnnn,sssss,aaaa,hhhhh
or
#adr,nnnnn rpm,sssss rpm,aaaa mA,hhhhh hours
Returns the current values:
nnnnn
: speed set point
sssss
: stand-by speed set point
aaaa
: current set point
hhhhh
: alert level for pump bearing maintenance
Complete cabinet only:
#adr
LEV
10<CR>[<LF>]
#adr,nnnnn,sssss,hhhhh,g,ccccc,eeeee,ddddd,pppp,qqqq
Returns current values:
nnnnn
: nominal speed set point (in rpm)
sssss
: stand-by speed set point (in rpm)
hhhhh
: alert level for pump bearing maintenance (in hours)
g
: regreasing counter
ccccc
: pump working time (in hours)
eeeee
: electronic working time (in hours)
ddddd
: start delay (max 14459 s, that is 240 mn 59 s)
pppp
: time to venting (max 3599 s, that is 59 mn 59 s)
qqqq
: venting time (max 3599 s, that is 59 mn 59 s)
Returns the strings sent with the identification sub-strings
#adr
LNG
<CR>[<LF>]
#adr,ok
CIT>
Allows the parameters returned by the DLR, LEV and SPD commands
to be identified with sub-strings.
Also generates the «CIT>» prompt each time a <CR> character is
received.
LNG
Syntax
Result
See also:
SHT