![Alcatel Vacuum Technology Adixen ATP 100 Скачать руководство пользователя страница 68](http://html.mh-extra.com/html/alcatel-vacuum-technology/adixen-atp-100/adixen-atp-100_user-manual_2891737068.webp)
B 130
Functions of integrable controllers
Brick and Board type
Alcatel Vacuum Technology France - ATP User’s Manual
1/4
GB 00421 - Edition 03 - March 00
Using the switch S3 :
Factory configuration :
9600 Bauds, WITHOUT
Parity, 8 bits, 1 Stop bit,
RS232.
Serial link
RS 232/485
Setting
Operation
Rotation speed setting
Refer to the Commands details sheet (
B 120
).
The potentiometer R15 (
See folio 4
) has to be set
to a position such as not to exceed the 1/5 of its
full excursion to allow the speed programming from the
serial link.
J4 Serial port
1
2
3
4
5
6
7
8
9
GND
V(+)
DTR
TD
RD
DSR
V(-)
120
Ω
(Internal)
8
7
6
5
4
3
2
1
ON
1
2
3
4
5
GND
Screen
Connect 7 to 8
if load impedance
necessary between A et B
B (or
V-)
A (or
V+)
6
7
8
9
RS 485 Link
1
2
3
4
5
GND
RS 232 Link
Screen
TD
(Terminal or computer)
RD
(Terminal or computer)
6
7
8
9
Switch
ON
OFF
Number of STOP bits
1
1 bit
2 bits
Data length
2
7 bits
8 bits
Parity (Valid if switch 4 is off)
3
ODD
EVEN
4
SANS
AVEC
Type of link
8
RS 485
RS 232
Switch
5
6
Transmission speed
9600
ON
ON
4800
OFF
ON
2400
ON
OFF
1200
OFF
OFF
Warning
Electronic boards have been designed to comply
with the electrical safety and electromagnetic
compatibility rules but they cannot be the purpose of
a certification as soon as they are sold alone.
It is the user responsibility to ensure the externals shields
which are necessary to be in compliance with ECM and
electrical safety rules