B 70
ACT 200 T and ACT 600 T
controller start-up
Alcatel Vacuum Technology France - ATP User’s Manual
2/4
GB 00415 - Edition 02 - Feburary 99
Configuring the controller
DISPLAY
SETUP
SETUP
RUNNING IN
Enter the access code
Configure the 0-10V output
Modify the STANDBY speed
Give the authorization to restart
the pump after a power cut
Activate or disactivate the buzzer
Select the temperature
measurement unit
Set the serial link parameters
Modify the time before starting
up the pump
Modify the time before opening the
venting valve
Modify the venting valve opening
time
Program the maximum operating
time before regreasing the
bearings
Ball bearings life time counter
Modify the access code
Enter the access code and
validate
SPEED: 6.75V = 27000rpm
I motor:
ATP 80/100: 2.5A / 10V
ATP 150/400: 5.1A / 10V
θ
.PUMP: 0.1V per 1°C
θ
.CONT: 0.1V per 1°C
6000 to 27000 rpm
YES or NO
ON or OFF
°C or °F
RS232/RS485/NETWORK
(see sub-menus folio 3 & 4)
0 to 240mn 59s
0 to 59mn 59s
0 to 59mn 59s
M=0 (1000 to 15000 h)
M=1 (1000 to 30000 h)
M=2 (1000 to 45000 h)
(3 times the maintenance time)
(see E 80)
0 to 50000 hours
0 to 65535
0
Speed
12000
NO
ON
°C
RS232
0mn 0s
0mn 1s
0mn 1s
M=0
5000h
0 h
0
ACCESS CODE
SET ANALOG-OUT
STANDBY SPEED
AUTO-STARTING
BUZZER
TEMPERATURE UNIT
SET SERIAL LINK
SET START DELAY
TIME TO VENTING
VENTING TIME
MAINTENANCE
TIME BEARING
NEW CODING
Factor
y
configuration
ENTER