E 20
ATP 80 and ATP 100
Pumps lubrication
Alcatel Vacuum Technology France - ATP User’s Manual
1/2
GB 00438 - Edition 01 - March 97
The first lubrication required for the correct operation of ATP
pumps is performed in the factory. Subsequent lubrications
should be performed according to the procedure below and
according to a frequency defined as a function of processes
used (see scales
D 30
).
Only use the ALCATEL grease contained in the lubrication
syringe
(refer to the maintenance component references
F 10
).
Avoid introducing foreign matter into the pump
during these operations. Lubrication must be
performed with the pump switched off.
PUMPING CELL
Lubrication of bearing on
the pumping cell side:
black jumper.
Lubrication of bearing
opposite the pumping
cell side:
red jumper.
INLET
EXHAUST
The ATP contains two bearings which must both be
reloaded with grease at each relubrication period.
The syringe is equipped with two jumpers with different
widths and colours (red and black)
used to proportion
the grease accurately for each pump bearing.
Use of the lubrication
syringe