B 100
ACT 1000 T controller start-up
Alcatel Vacuum Technology France - ATP User’s Manual
2/4
GB 00418 - Edition 02 - February 98
Programming
the parameters
* At power-up, the storage mode disappears automatically.
PARAMETERS
Modify the date
Modify the STANDBY speed
Modify the time before starting
up the pump
Modify the maximum bearing
operating time
Modify the RS232 serial link
Configure the 0-10V output
Return to the factory
configuration (Alcatel reserved)
Modify the time before
opening the air inlet valve
Modify the air inlet valve
opening time
Change to storage mode
(the timer is blocked).
6000 to 27000 tr/mn
0 to 59mn 59s
M=0 (5000 to 15000 h)
M=1 (5000 to 30000 h)
M=2 (5000 to 45000 h)
(see details folio 3)
θ
.PUMP
0.1V per 1°C
6,75V = 27000 tr/mn
θ
.CONV
0.1V per 1°C
Current 10V for 15000mA
0 to 59mn 59s
0 to 59mn 59S
ON / OFF
12000
0mn 0s
0
5000 h
Θ
TEMP.
1s
1s
ON*
DATE
SPEED
START
DELAY
BEARING
MAINT
LIMIT
RS 232
AN.OUT
Θ
1 PUMP
SPEED
Θ
CONV
I.MOTOR
RESET
CONTROLLER
BEARING
AIR INLET
DELAY
OPEN
STOCK
Factor
y
configuration
ENTER
ENTER
ENTER
ENTER