10.8 Operating Parameters
Some operating parameters are software options and can be user-configured
with the function ´Config´ (see manual 6.10.13.2).
Mains frequency noise suppression 60Hz instead of 50Hz
Config: F-------
Clearing all measured values on start of a measurement
Config: -C------
Ring memory (overwriting of old data when full)
Config: --R-----
Year number output in date with 4-digits instead of 2-digits
Config: ---D----
Deactivating the alarm value/alarm printout
Config: ----A---
Switching off the signal transmitter
Config: -----S--
Date-time in Excel table format
´dd.mm.yy hh:mm:ss´
Config: ------E-
The following parameters can be used to check the device function:
The channel number can be configured, e.g. for
hiding upper channels (see manual 6.10.13.1):
Out of 40 or 36, respectively, 5 channels are activated:
Meas.Chann:40
active:05
Sensor supply voltage 12.3V = mains operation:
SensorVoltage: 12.3 V
Cold junction temperature = socket temperature:
CJ Temperature: 25.4°C
11. OUTPUT MODULES
The data logger ALMEMO
®
2590-9 has two
output sockets, A1 and A2, to allow an output
of the measured values as analog or digital
values or as alarm signals. Furthermore, it is
possible to trigger various functions through
trigger pulses. To suit all possibilities, while
keeping the hardware requirements at a
minimum, all necessary interfaces have been
integrated into the ALMEMO
®
output
connectors. The output modules, like the
sensors, are automatically detected and
displayed in the menu ´Output Modules´ so
that, generally, no programming is required.
11.1 Data Cable
The serial interface can be used to output cyclic data logs, all function values
of the measuring menus, as well as the whole programming of the sensors and
the device to a printer or computer. The ALMEMO
®
data cables and the
connection to the instruments are described in the manual section 5.2. Other
modules for networking the instruments follow in the manual section 5.3.
All available interface modules are connected to the socket A1(3), exception:
network cables ZA1999NK to network a further instrum. are connected to A2.
62
ALMEMO
®
2590-9
Output Modules
* OUTPUT MODULES
*
Socket A1:
DK Data Cable
0: RS232
Socket A2:
EA Trigger Alarm
2: Rx int. allocated
Relay: 01------
Analog Chann: 00
Analog Val: +32500